Abstract
<jats:p>ABSTRACT Background: Metallic implants such as orthopedic screws, prostheses, and dental hardware produce beam-hardening, photon-starvation, and streak artifacts that degrade computed tomography (CT) image quality, and the metal artifact reduction (MAR) methods developed to mitigate them require objective, reproducible benchmarking. Purpose: Objective evaluation of MAR algorithms in CT is hindered by the absence of phantoms that simultaneously provide anatomically realistic backgrounds, embedded implants of known geometry, and controllable, ground-truth—referenced artifact intensity. We present a dual-filament, voxel-level three-dimensional (3D) printing method that fulfills these requirements and demonstrate its capabilities on a clinically representative cervical spine case with embedded orthopedic spinal screws. Methods: The proposed method extends the PixelPrint framework, a fused-deposition-modeling (FDM) workflow that converts clinical Digital Imaging and Communications in Medicine (DICOM) data directly into 3D-printer Geometric code (G-code) without intermediate segmentation or surface meshing, to interleaved, voxel-level deposition of two filaments: a calcium-doped polylactic acid (PLA) for soft tissue and bone, and a higher-attenuation metal-doped PLA for metallic implants. For demonstration, anonymized DICOM data of a healthy cervical spine were used to design and fabricate three matched phantoms, each with six embedded spinal screws at C4—C6: a 0% metal-infill ground-truth phantom, a 50% medium-metal-infill phantom, and an 85% high-metal-infill phantom. All phantoms were scanned on a clinical spectral CT system at 120 kVp and 1000 mAs, reconstructed at 0.67 mm slice thickness with virtual monoenergetic imaging (VMI) across 50—190 keV. Method performance was characterized by region of interest (ROI)-based Hounsfield Unit (HU) agreement with the source patient data and by the noise-independent Gumbel-distribution p-index metric. Results: The dual-filament method reproduced patient anatomy, soft-tissue contrast, and screw geometry with high fidelity. ROI HU values agreed with patient data within ±25 HU for soft tissue and trabecular bone; cortical regions were underestimated owing to the current ceiling of the calcium-doped PLA used in this study. The tunable-artifact behavior was quantified as follows: the Gumbel location parameter μ scaled monotonically from 46.7 HU (no-metal background) to 57.1 HU (50% infill) to 90.5 HU (85% infill) for the VMI 70 keV with standard filter. High-keV VMI reconstructions substantially reduced streak and beam-hardening artifacts while preserving anatomic detail. Conclusions: The proposed dual-filament, voxel-level PixelPrint method enables the fabrication of patient-specific, multi-material CT phantoms with embedded metallic implants and controllable, ground-truth—referenced artifact intensity. Although demonstrated here in a single cervical-spine case, the workflow is anatomy- and implant-agnostic by construction and could in principle be adapted to other musculoskeletal sites (e.g., knee, hip, dental) and implant materials, providing a reproducible methodological foundation for benchmarking MAR algorithms, characterizing spectral CT performance, and validating emerging photon-counting detector systems. Keywords: 3D printing methodology; fused deposition modeling; voxel-level multi-material printing; spectral computed tomography; metal artifact reduction; phantom design; orthopedic implants; dual filament; PixelPrint.</jats:p>