Deprecated: Function curl_close() is deprecated since 8.5, as it has no effect since PHP 8.0 in /home/u483256323/domains/poorvam.com/public_html/subdomains/pore/includes/api.php on line 184
Back to Search View Original Cite This Article

Abstract

<jats:p>In this study, we performed radiation shielding calculations for the diagnostic X-ray equipment at Rekha Med Hospital. The lead thickness required for radiation protection, as determined by the methodology outlined in the National Council on Radiation Protection and Measurements (NCRP) Report No. 147, was approximately 2 mm. Air kerma measurements were taken outside the X-ray room to validate the shielding effectiveness with a 2 mm thick lead barrier installed along the walls. The results confirmed that the air kerma dose did not exceed the regulatory dose limit of 0.5 mSv/H in uncontrolled areas, indicating sufficient shielding.</jats:p>

Show More

Keywords

radiation shielding xray lead protection

Related Articles


Deprecated: Function curl_close() is deprecated since 8.5, as it has no effect since PHP 8.0 in /home/u483256323/domains/poorvam.com/public_html/subdomains/pore/includes/api.php on line 76
PORE

About

Connect